![KCI V.A.C. ATS Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/kci/v-a-c-ats/v-a-c-ats_operation-and-maintenance-manual_3787715029.webp)
V.A.C.
ATS Operations and Maintenance Manual
25
Firmware Upgrade
WARNING!
Starting this procedure will prevent the therapy unit from functioning
until upgrade is successfully completed. Ensure that the correct Download software is
installed on your PC and that you have the correct product software update to
complete the process.
EQUIPMENT REQUIRED
-
PC with serial com port available, (preferably COM1), and either
Windows 95, 98 or NT operating system.
-
Serial upgrade cable, (MiniVAC type is acceptable) – M6243830
-
PC Download software installed, available from UK Engineering
-
V.A.C ATS Software upgrade file
SETTING UP
1.
Install the PC Download software on the PC.
2.
Connect the serial upgrade cable to a free serial port. If the port is not COM1,
refer to the ReadMe.txt file for the correct command syntax to use other serial
ports.
PROCEDURE
1.
Switch on the therapy unit and press Utilities the Service. The following Access
Code screen will appear.
2.
Enter the software upgrade access code 524911 and press OK. The screen will go
blank.
3.
RESTRICTED USE
Certain pages of this manual are designated
“Restricted Use Only.” Information contained on
those designated pages may be used solely to
facilitate the proper maintenance and repair of
KCI products identified herein. If these terms are
unacceptable, immediately remove those
designated pages and return them to KCI, Attn:
Legal Dept. Refer to qualified service personnel
who are authorized by KCI for proper
maintenance and repair