V.A.C.
ATS Operations and Maintenance Manual
17
4.0
CANISTER HOUSING REPLACEMENT
4.1
Remove the Front Housing,
and the Secondary Filter.
4.2
Disconnect the Battery.
4.3
Undo the 3 screws that
secure the Canister Holder
to the Rear Housing.
4.4
Disconnect the large and
small tubes from the top of
the Bellows Lever.
4.5
Disconnect the ribbon cable
from the Canister Board.
4.6
Reverse the above
procedure to install the new
Canister Housing Assembly.
NOTE:
When re-connecting the Large and
Small tubing, ensure that the elbow
connectors are positioned as shown
opposite.
The large elbow is designed to locate on
a small notch at the front.
5.0
BATTERY REPLACEMENT
5.1
Remove the Front Housing.
5.2
Disconnect the tubing from the right side of
the manifold.
5.3
Disconnect the battery from the charger
board.
5.4
Unscrew the Battery Retaining Clip and
withdraw the battery from the unit.
5.5
Remove the battery from the metal ‘clip’
and replace with a new pack.
5.6
Ensure that the leads exit the Battery
Retaining clip as shown opposite.
5.7
Reverse the above procedure to install the
new pack.
NOTE:
The battery’s serial number and manufactured date is recorded on the side of
the battery pack.
Add the battery manufacturing date to the ‘Battery History’ label,
and fix to the underside of the Therapy unit.
4.2
4.3
4.6
RESTRICTED USE
Certain pages of this manual are designated
“Restricted Use Only.” Information contained on
those designated pages may be used solely to
facilitate the proper maintenance and repair of
KCI products identified herein. If these terms are
unacceptable, immediately remove those
designated pages and return them to KCI, Attn:
Legal Dept. Refer to qualified service personnel
who are authorized by KCI for proper
maintenance and repair