Table 2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity to Interference
The tanning device is intended for use in the electromagnetic environment specified below. The studio operator
of the tanning device should assure that it is used in such an environment.
Immunity tests
IEC 60601-
Test level
Compliance level
Electromagnetic environment
- Guidance
Electrostatic discharge accord-
ing to IEC 61000-4-2
± 8 kV
contact discharge
± 15 kV
air discharge
+/- 8 kV
contact discharge
+/- 15 kV
air discharge
Floors should be made of wood or concrete
or be covered with ceramic tiles. If floors are
covered with synthetic material, the relative
humidity needs to be at least 30 %.
Fast transient electrical distur-
bances/bursts according to IEC
61000-4-4
± 2 kV
for power-supply lines
± 1 kV
for input and output lines
± 2 kV
for power-supply lines
± 1 kV
for input and output lines
Mains power quality should be that of a typi-
cal industrial or hospital environment.
Surge voltages (surges)
according to IEC 61000-4-5
± 1 kV
differential mode voltage
± 2 kV
common mode voltage
± 1 kV
differential mode voltage
± 2 kV
common mode voltage
Mains power quality should be that of a typi-
cal industrial or hospital environment.
Voltage dips, short interruptions
and voltage variations on pow-
er-supply input lines according
to IEC 61000-4-11
Dips
0% for 1 cycle
70% for 25/30 cycles
Interrupts
0% for 250/300 cycles
Dips
0% for 1 cycle
70% for 25/30 cycles
Interrupts
0% for 250/300 cycles
Mains power quality should be that of a typi-
cal industrial or hospital environment.
If the studio operator of the tanning device
requires continuous operation of the device
even in the event of interruptions of the pow-
er supply, it is recommended that the tan-
ning device is powered by an uninterrupted
power supply unit.
Power-frequency (50/60 Hz)
magnetic field according to IEC
61000-4-8
30 A/m
30 A/m
Power-frequency magnetic fields should be
at levels characteristic of a typical location in
a typical industrial or hospital environment.
O p e r a t i n g I n s t r u c t i o n s
|
KBL 4800 alpha
16
Содержание 4800 alpha
Страница 1: ...OPERATING INSTRUCTIONS TANNING DEVICE ...
Страница 2: ......
Страница 57: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL 4800 alpha 57 ...
Страница 58: ...Remarks O p e r a t i n g I n s t r u c t i o n s KBL 4800 alpha 58 ...
Страница 59: ......
Страница 60: ...Ringstraße 24 26 56307 Dernbach Germany fon 49 0 26 89 94 26 0 fax 49 0 26 89 94 26 66 ...