Chapter 7 Starting Standalone Manual Calibration
Kaye HTR-420 User Manual
33
Starting Standalone Manual Calibration
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in serious
injuries.
•
Never touch the metal block, the dry block insert, external reference probe, or the test specimen at
temperatures above 35°C.
•
Allow the calibrator to cool before you remove the test specimen/external reference probe, change
the dry block insert or switch off the calibrator.
Before the testing process
Before the start of the testing process, check whether:
The instructions regarding the installation site and the operational position have been complied with
The electrical connections have been made correctly.
The electrical connection has been grounded – otherwise the HTR-420 will not heat nor cool.
The correct measuring insert has been selected for the test task.
The test sensors are securely fixed in the calibrator.
The calibrator has sufficient structural stability.
Starting the testing process:
IMPORTANT! Parameter fields disabled!
During the testing process, all the parameter fields, except for the Start / Stop slider and the
alarm signal, are disabled.
Push the button of the start / stop slider completely to the right.
The testing process is starts. Information on the testing process and the temperature
progression displays in the window.