Chapter 2. Safety Information
Kaye HTR-420 User Manual
13
IMPORTANT NOTE!
The thermocouple holder may be hot after use. Wait for the HTR-420 to cool to a safe
temperature before touching or removing the holder.
IMPORTANT NOTE!
Keep in mind after transport or storage!
After transport, storage or long periods of non-use, moisture can seep into the heating
elements (magnesium oxide). For drying, the calibrator must be slowly heated up. During
this process, the calibrator has not yet reached the required insulation voltage for
protection class I.
For drying the heating elements, heat the calibrator for at least 15 min to 120 °C.
Further warnings that are specifically relevant to individual operating procedures or activities can
be found at the beginning of the relevant sections of this user manual.