Chapter 4. Getting Started
Kaye HTR-420 User Manual
17
4.4
HTR-420 Startup
Prior to powering up the HTR-420 perform the following task;
Push the dry block insert down to the stop. Connect the External Reference Probe to the
connector marked “ext Ref” on the front panel and insert into the dry block using the
4.8mm hole sized specifically for the 4.5mm Reference Control Probe.
Power On the HTR-420 by utilizing the Power Switch on the front panel.
Once powered on, the Kaye HTR-420 splash screen appears for approximately thirty seconds,
followed briefly by the model and firmware version, before presenting the
Remote Control
home
screen.
Since the HTR-420 is predominately used for automatic calibration with Kaye validation
equipment, the unit automatically boots into the
Remote Control
screen with the following default
settings:
•
Mode
Remote Control mode
•
Function:
Dry Block
•
Control Temp:
External Reference Probe
•
Temp Controller
OFF
•
Language Selection:
English
•
Date and Time:
MM.DD.YYYY and 24h
•
Temperature Units:
Celsius
•
Decimal Places:
0.00