7 Reprocessing steps in accordance with ISO 17664 | 7.5 Care products and systems - Servicing
20 / 28
7.5 Care products and systems - Servicing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.
Note
KaVo guarantees the proper function of KaVo products only if the care
products listed as accessories are used, since these were tested for proper
use on our products.
7.5.1 Care with KaVo Spray
Note
Head can be serviced alone or attached to a shank.
KaVo recommends servicing the product after each use, i.e. after each auto-
matic cleaning and before each sterilisation.
▶
Remove the tool from the medical device.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.
▶
Press the spray key for 1 to 2 seconds.
Care of chucking system
KaVo recommends cleaning and servicing the chuck system once a week.
▶
Remove the tool from the medical device.
▶
Position the tip of the spray nipple in the opening, and apply the spray.
▶
Press the spray key for 1 to 2 seconds.
7.5.2 Servicing with KaVo QUATTROcare PLUS
Note
Head can be serviced alone or attached to a shank.
Cleaning and servicing device with expansion pressure for internal cleaning of
inorganic residues and optimum care.
(no validated cleaning of the interior according to German Robert Koch Institute
(RKI) requirements)