7 Reprocessing steps in accordance with ISO 17664 | 7.3 Manual reprocessing
18 / 28
▶
Brush the spray clip and spray tube under running tap water water for at
least 20 seconds using a medium-hard toothbrush.
For validated internal cleaning of the spray clip and spray tube in the cleaning
and disinfecting device, preliminary non-fixing cleaning is required.
7.3 Manual reprocessing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
Clean it in a washer disinfector only.
Note
The spray clip for internal cooling may be additionally cleaned in the ultra-
sonic cleaner.
7.3.1 Manual internal and external cleaning and internal
and external disinfection
Not applicable.
7.3.2 Manual drying
▶
Blow off the outside and inside with compressed air until water drops are no
longer visible.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
See also:
2
7.5 Care products and systems - Servicing, Page 20
7.4 Automated reprocessing
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.