Instructions for use SURGmatic S201 L - 1.009.0470 SURGmatic S201 C - 1.009.1100
4 Start up and shut down | 4.1 Checking the amount of water
10 / 28
4 Start up and shut down
WARNING
Hazard from contaminated products.
Infection hazard for care provider and patient.
▶
Prior to initial commissioning and after each use, reprocess the product and
accessories.
See also:
2
7 Reprocessing steps in accordance with ISO 17664, Page 17
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
▶
Reprocess and sterilise the product and accessories before disposal.
See also:
2
7 Reprocessing steps in accordance with ISO 17664, Page 17
4.1 Checking the amount of water
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Thermal damage to the dental pulp.
▶
Adjust the water amount for the spray cooling to a minimum of 50 ml/min!
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶
Check the spray water channels and clean the spray tube with the nozzle
needle (
Mat. no. 0.410.0931
) according to need.
▶
Switch off spray-air and spray-water supply on the treatment device.