![KaVo SURGmatic S201 C Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/kavo/surgmatic-s201-c/surgmatic-s201-c_instructions-for-use-manual_4007762012.webp)
Instructions for use SURGmatic S201 L - 1.009.0470 SURGmatic S201 C - 1.009.1100
5 Operation | 5.1 Inserting the head
12 / 28
5 Operation
5.1 Inserting the head
WARNING
Detachment of the medical device during treatment.
If the head is not properly locked in place, it can become detached during
treatment.
▶
Do not mount or remove the head while it is rotating. Before each treat-
ment, check that the head is firmly seated and that the clamping ring is
tight.
CAUTION
It is not permissible to combine the unit with other heads / bases.
Risk of injury.
▶
The SURGmatic S201 L/ C contra-angle handpiece consists of a base and a
head and must not be combined with other heads or bases.
Note
The head of the SURGmatic S201 L / C contra-angle handpiece should be
taken off the base for reprocessing purposes only.
▶
Rotate the clamping ring in the direction of the arrow until it hits the stop
and hold it there.
▶
Insert the head to the stop. Make sure that the catches engage properly.
▶
Rotate the clamping ring in the direction of the arrow (-> close) and tighten
it.
▶
Attach the spray clip.
▶
Make sure that the spray clip is securely affixed.
5.2 Pulling off the head
▶
Pull the spray clip off.
▶
Rotate the clamping ring in the direction of the arrow until it hits the stop
and hold it there.
▶
Pull off the medical device.
▶
Release the clamping ring.