Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005
7 Reprocessing steps in accordance with EN ISO 17664 | 7.5 Care products and systems - Servicing
18 / 24
▶
Plug the product onto the cannula, and press the spray button for one
second.
Servicing of the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a week.
▶
Remove the cutter or grinder, place the spray nipple tip in the opening and
spray.
Note
Carry out servicing according to instructions in the section "Care with KaVo
Spray".
7.5.2 Servicing with KaVo QUATTROcare PLUS
Cleaning and servicing device with expansion pressure for internal cleaning of
inorganic residues and optimum care.
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Remove the tool from the medical device.
▶
Service the product in the QUATTROcare PLUS.
See also:
2
Instructions for use KaVo QUATTROcare PLUS
Servicing the clamping chuck
KaVo recommends servicing the chuck system once a week using the chuck
servicing program integrated in the device.
Note
Handpieces must be taken off the service couplings before the chuck service
can be started and performed.
▶
Close the front door and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecut-
ively.
ð
The device is in chuck service mode.
▶
Remove the service coupling of the chuck from the side hatch of the QUAT-
TROcare PLUS and attach it to coupling service point four, on the far right.
A MULTIflex adaptor must be mounted there.
▶
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
▶
Press the button marked with the chuck service symbol.