Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005
7 Reprocessing steps in accordance with EN ISO 17664 | 7.7 Sterilisation
20 / 24
7.7 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with ISO 17665-1
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶
Before each sterilisation cycle, service the medical device with KaVo care
products.
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶
Immediately remove the product from the steam steriliser after the steril-
isation cycle!
The KaVo medical device has a maximum temperature resistance up to 138
℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilisation processes:
▪ Steriliser with triple pre-vacuum:
- at least 3 minutes at 135 °C (275 °F)
– Drying time: 16 min.
▪ Steriliser using the gravity method:
- at least 10 minutes at 135°C (275 °F)
– Drying time: 30 min.
▶
Use according to the manufacturer's Instructions for Use.
7.8 Storage
Prepared products must be stored appropriately in a dry, dark, cool room such
that they are protected from germs (as far as possible) and dust.
▶
Comply with the expiry date of the sterilised items.