Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005
7 Reprocessing steps in accordance with EN ISO 17664 | 7.1 Preparation at the site of use
7 Reprocessing steps in accordance with EN ISO 17664
7.1 Preparation at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶
Take suitable personal protective measures.
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
▶
Remove the tool from the medical device.
▶
Remove all residual cement, composite or blood immediately.
▶
Reprocess the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to reprocessing.
▶
Do not place in solutions or similar substances.
7.2 Non-fixing preliminary cleaning of the spray tube
Accessories required:
▪ Demineralised water 30 °C ± 2 °C (86 °F ± 3.6 °F)
▪ Nozzle pin
▪ Brush, e.g. medium-hard toothbrush
▪ Disposable syringe
▶
Check the patency of the spray tube and clean it using the nozzle needle
(
Mat. no. 0.410.0931
).
▶
Rinse the spray tube with at least 20 ml demineralised water using a dis-
posable syringe.
▶
If the spray tube is not patent after the manual rinsing procedure, the med-
ical device must be replaced.
Brush the spray tube and hose under running tap water water for at least 20
seconds using a medium-hard toothbrush.
The non-fixing preliminary cleaning is a central constituent and must be carried
out prior to the automatic reprocessing.
In the cleaning and disinfecting device, validated internal cleaning of the spray
tube necessitates preliminary non-fixing cleaning.
15 / 24