Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005
4 Start up and shut down | 4.1 Checking the amount of water
4 Start up and shut down
WARNING
Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶
Before first startup and after each use, reprocess and sterilise the medical
device and accessories accordingly.
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
▶
Before disposal, reprocess and sterilise the product and accessories appro-
priately.
4.1 Checking the amount of water
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Thermal damage to the dental pulp.
▶
Adjust the water amount for the spray cooling to a minimum of 50 ml/min!
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶
Check the spray water channels and clean the spray tube with the nozzle
needle (
Mat. no. 0.410.0931
) according to need.
▶
Switch off spray-air and spray-water supply on the treatment device.
▶
Cooling the drill bit or bur via external supply.
▶
During surgical interventions, comply with the necessary precautions re-
garding cooling.
▶
Use physiological, sterile cooling fluid.
11 / 24