5 Operation | 5.1 Attach the medical device
12 / 24
5 Operation
5.1 Attach the medical device
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked in place can become disconnected
from the motor coupling and fall off.
▶
Carefully pull on the medical device before each treatment to ensure that it
is securely locked onto the motor coupling.
CAUTION
Connect to the drive motor.
Handpiece blocked.
▶
Only start the handpiece when the chuck is closed.
NOTICE
Removing and attaching the straight or contra-angle handpiece while
the drive motor is rotating.
Damage to the driver.
▶
Never attach or remove the straight or contra-angle handpiece while the
drive motor is rotating.
▶
Lightly spray O-rings on motor coupling with KaVo Spray.
▶
Place the medical device on the motor clutch and lock it into place. With the
SURGmatic S11 L, the latch must lock into place audibly.
▶
Pull on the medical device to make sure that it is securely affixed to the
coupling.
5.2 Remove the medical device
▶
Unlock the medical device from the motor coupling by twisting it slightly
and then pulling it along its axis.
5.3 Insert the handpiece or contra-angle handpiece drill
bit.
Note
Only use handpiece or contra-angle handpiece burs that correspond to EN
ISO 1797 type 1 and type 2, are made of steel or hard metal and meet the
following criteria:
Shaft diameter: Ø 2.334 to Ø 2.350 mm
Contra-angle burs with bur stop:
- Shaft clamping length: at least 12 mm
- Overall length: max. 22 mm
Handpiece burs without bur stop:
- Shaft clamping length: at least 30 mm
- Overall length: max. 44.5 mm