5 Operation | 5.1 Attaching the medical device
5 Operation
5.1 Attaching the medical device
WARNING
Release of the medical device during treatment.
A medical device that is not properly locked in place can release from the mo-
tor coupling and fall off.
▶
Carefully pull on it before each treatment to ensure that the medical device
is securely locked on the motor coupling.
NOTICE
Removing and attaching the contra-angle handpiece while the drive
motor is rotating.
Damage to the driver.
▶
Never attach or remove the contra-angle handpiece while the drive motor
is rotating.
NOTICE
Pressing the foot switch while attaching or detaching the medical
device.
Property damage to the medical device.
▶
Do not connect or remove the medical device while pressing the foot
switch.
▶
Lightly spray O-rings on motor coupling with KaVo Spray.
▶
Place the medical device on the motor coupling and lock it into place.
▶
Pull on the medical device to make sure that it is securely affixed to the
coupling.
5.2 Removing the medical device
▶
Unlock the medical device from the motor coupling by twisting it slightly
and then pulling it along its axis.
5.3 Inserting the prophy cups or polishing brushes
Note
Only use prophy cups or polishing brushes in accordance with DIN EN ISO
13295 for support (spindle) type 5.
WARNING
Use of impermissible inserts.
Injury to the patient or damage to the medical device.
▶
Observe the instructions for use, and use the insert properly.