Instructions for use SMARTmatic PROPHY S31 - 1.011.6760 SMARTmatic PROPHY S31 K - 1.011.6761 SMARTmatic
PROPHY S33 - 1.011.6800
5 Operation | 5.4 Removing prophy cups or polishing brushes
14 / 26
CAUTION
Injury from using worn inserts.
The inserts can fall out during treatment and harm the patient.
▶
Never use insert with worn thread.
CAUTION
Danger of injury from insert.
Infections or cuts.
▶
Wear gloves or fingerstalls.
CAUTION
Risk of burn injury from hot handpiece head.
If prophylaxis cups touch the handpiece head, the head may heat up due to
friction. Burn injuries in the mouth may be caused if the handpiece overheats.
▶
Do not use prophylaxis cups that touch against the handpiece head.
▶
Block the catch pin of the contra-angle handpiece with the enclosed holder.
▶
Screw or push the insert into/onto the head.
▶
Remove the holder.
▶
Check that the insert is firmly seated.
5.4 Removing prophy cups or polishing brushes
WARNING
Hazard from rotating insert.
Lacerations.
▶
Do not touch rotating insert!
▶
Remove the insert from the contra-angle handpiece after treatment to
avoid injury and infection during storage.
▶
Block the catch pin of the contra-angle handpiece with the enclosed holder.
▶
Unscrew and/or take off the insert from the head.