7 Reprocessing steps in accordance with ISO 17664 | 7.3 Disinfection

Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter. Press the spray button three times for 2
seconds each time. Remove the medical device from the spray attachment
and let the cleanser act for one minute.

Afterwards, rinse for 3-5 seconds with KaVo DRYspray.

See also:

2

KaVo CLEANspray / KaVo DRYspray Instructions for Use

Note

KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only
available in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxembourg,
Belgium, Netherlands, United Kingdom, Denmark, Sweden, Finland and Nor-
way.
In other countries interior cleaning can only be carried out with washer disin-
fectors in accordance with EN ISO 15883-1.

7.2.4 Automated internal cleaning

KaVo recommends washer disinfectors from Miele in accordance with EN ISO
15883-1 that are operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD programme, the cleaning
agent neodisher® MediClean and the neutralisation agent neodisher® Z.

For programme settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the washer disinfector.

In order to prevent damage to the medical device due to residual fluid,
make sure that the inside and outside of the device is dry after the end of
the cycle. Remove any residual liquids from the interior and exterior of the
medical device using compressed air.

Immediately after drying, lubricate the medical device with care agents
from the KaVo care system.

The drying procedure is normally part of the cleaning program of the thermod-
isinfector.

Note

Please comply with the instructions for use of the washer disinfector.

7.3 Disinfection

 WARNING

Incomplete disinfection.

Infection hazard

As a matter of principle, KaVo recommends carrying out a final disinfection
of the unpackaged item in a steriliser if complete disinfection cannot be
guaranteed without this measure.

NOTICE

Using the disinfectant bath or chlorine-containing disinfectants.

Malfunction and material damage.

Do not disinfect the device in the disinfection bath or with chlorine-contain-
ing disinfectants.

17 / 26

Содержание SMARTmatic PROPHY S31

Страница 1: ...Instructions for use SMARTmatic PROPHY S31 1 011 6760 SMARTmatic PROPHY S31 K 1 011 6761 SMARTmatic PROPHY S33 1 011 6800 ...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Страница 3: ...ings 15 7 Reprocessing steps in accordance with ISO 17664 16 7 1 Preparations at the site of use 16 7 2 Cleaning 16 7 2 1 Manual external cleaning 16 7 2 2 Automated external cleaning 16 7 2 3 Manual internal cleaning 16 7 2 4 Automated internal cleaning 17 7 3 Disinfection 17 7 3 1 Manual external disinfection 18 7 3 2 Manual internal disinfection 18 7 3 3 Automated external and internal disinfec...

Страница 4: ...have any questions or complaints please contact the KaVo Technical ser vice 49 0 7351 56 1000 service instrumente kavokerr com Target group This document is intended for dentists and their assistants The startup section is also intended for service technicians General marks and symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Action requ...

Страница 5: ...essure Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material damage The warning notes are designated as shown below DANGER In cases which if not prevented directly lead to death or severe in jury WARNING In cases which if not prevented co...

Страница 6: ...ely repro cessed or sterilised In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air Use gloves or finger guard whenever you test insert and remove the tool 2 2 Technical condition A damaged device or components could injure patients users and third parties Only operate devices or components if they are undamaged on the outside ...

Страница 7: ...h a treatment centre released by KaVo Comply with the Instructions for Use of the treatment centre 2 4 Qualification of personnel Application of the product by users without the appropriate medical training could injure the patients the users or third parties Make sure that the user has read and understood the instructions for use Only employ the device if the user has the appropriate medical trai...

Страница 8: ...SMARTmatic PROPHY S33 1 011 6800 2 Safety 2 5 Service and repair 8 26 KaVo recommends specifying in house service intervals where the medical device is brought to a professional shop for cleaning servicing and a func tion check Defined the service interval depending on the frequency of use ...

Страница 9: ... PROPHY S31 K 1 011 6761 SMARTmatic PROPHY S33 1 011 6800 3 Description of the product 2 5 Service and repair 3 Description of the product SMARTmatic PROPHY S31 Mat No 1 011 6760 SMARTmatic PROPHY S31 K Mat No 1 011 6761 SMARTmatic PROPHY S33 Mat No 1 011 6800 9 26 ...

Страница 10: ...evention regulations these Instructions for use According to these regulations the user is required to to only use equipment that is operating correctly adhere to the specified intended use to protect him or herself the patient and third parties from hazards and to prevent contamination from the product 3 2 Technical specifications for S31 and S31 K Drive speed max 20 000 rpm Speed transmission 8 ...

Страница 11: ...dentistry can be attached The contra angle handpiece S33 can be mounted on all INTRAmatic LUX mo tors and motors fitted with a connector in accordance with ISO 3964 DIN 13940 Note The support spindle for the snap on polishing attachments cannot be un screwed 3 4 Transportation and storage conditions NOTICE Startup after refrigerated storage Malfunction Prior to startup strongly refrigerated produc...

Страница 12: ...and accessories WARNING Dispose of the product in appropriate manner Infection hazard Reprocess and sterilise the product and accessories before disposal NOTICE Damage from soiled and moist cooling air Contaminated and moist cooling air can cause malfunctions Make sure that the supply of cooling air is dry clean and uncontaminated according to EN ISO 7494 2 NOTICE Damage to the contra angle handpi...

Страница 13: ...is rotating NOTICE Pressing the foot switch while attaching or detaching the medical device Property damage to the medical device Do not connect or remove the medical device while pressing the foot switch Lightly spray O rings on motor coupling with KaVo Spray Place the medical device on the motor coupling and lock it into place Pull on the medical device to make sure that it is securely affixed t...

Страница 14: ...the head may heat up due to friction Burn injuries in the mouth may be caused if the handpiece overheats Do not use prophylaxis cups that touch against the handpiece head Block the catch pin of the contra angle handpiece with the enclosed holder Screw or push the insert into onto the head Remove the holder Check that the insert is firmly seated 5 4 Removing prophy cups or polishing brushes WARNING...

Страница 15: ...ice is too hot while working Service the medical device When the speed drops or is uneven Service the medical device An O ring is missing on the motor coupling Replace O ring See also 2 Instructions for use of motor 6 1 Troubleshooting 6 1 1 Replacing the O rings CAUTION Hazard from improper care of the O rings Functional problem or complete failure of the medical device Do not use Vaseline or oth...

Страница 16: ...ess this medical device in an ultrasonic device Malfunction and material damage Clean manually or in a washer disinfector only 7 2 1 Manual external cleaning Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Brush e g medium hard toothbrush Brush off under flowing tap water 7 2 2 Automated external cleaning KaVo recommends washer disinfectors from Miele in accordance with EN ISO 15883 1 th...

Страница 17: ...ine cleaning agents The validations were conducted with the VARIO TD programme the cleaning agent neodisher MediClean and the neutralisation agent neodisher Z For programme settings as well as cleansers and disinfectants to be used please refer to the Instructions for Use of the washer disinfector In order to prevent damage to the medical device due to residual fluid make sure that the inside and ...

Страница 18: ...have been released by KaVo with respect to the compatibility of materials e g WL cid made by ALPRO Immediately after internal disinfection lubricate the KaVo medical device with care agents from the KaVo care system 7 3 3 Automated external and internal disinfection KaVo recommends washer disinfectors from Miele in accordance with EN ISO 15883 1 that are operated with alkaline cleaning agents The ...

Страница 19: ... of KaVo products only if the care products listed as accessories are used since these were tested for proper use on our products 7 5 1 Servicing with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Take screw in and or snap on polishing attachments off the contra angle handpiece Cover th...

Страница 20: ...achments off the contra angle handpiece Service the product See also 2 Instructions for use KaVo QUATTROcare 2104 2104A 7 5 4 Servicing with KaVo QUATTROcare PLUS Servicing and cleaning device with expansion pressure for the interior cleaning of inorganic residues and optimum care no validated cleaning of the interior according to German Robert Koch Institute RKI requirements KaVo recommends servi...

Страница 21: ...ely remove the product from the steam steriliser after the steril isation cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Select a suitable procedure depending on the available autoclave from the fol lowing sterilisation processes Steriliser with triple pre vacuum at least 3 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Steriliser using the gravity method at least...

Страница 22: ...nt stand 2151 0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Holder 1 004 0596 Spray head INTRA KaVo Spray 0 411 9911 INTRA service coupling 1 009 6143 Spindle 0 549 0602 Material summary Mat No Adapter INTRAmatic CLEANspray and DRYspray 1 007 1776 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spray 2112 A 0 411 9640 ROTAspray 2 2142 A 0 411 752...

Страница 23: ...hall not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo s instructions for use or other manu...

Страница 24: ......

Страница 25: ......

Страница 26: ...1 012 7185 Fk 20180219 3 en ...

Отзывы: