7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
16 / 26
7 Reprocessing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from non-sterile products.
There is a risk of infection from contaminated medical devices.
▶
Take suitable personal protective measures.
▶
Take screw-in and/or snap-on polishing attachments off the contra-angle
handpiece.
▶
Remove all residual cement, composite or blood immediately.
▶
Reprocess the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to reprocessing.
▶
Do not place in solutions or similar substance.
7.2 Cleaning
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
Clean manually or in a washer disinfector only.
7.2.1 Manual external cleaning
Accessories required:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush, e.g. medium-hard toothbrush
▶
Brush off under flowing tap water.
7.2.2 Automated external cleaning
KaVo recommends washer disinfectors from Miele in accordance with EN ISO
15883-1 that are operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD programme, the cleaning
agent neodisher® MediClean and the neutralisation agent neodisher® Z.
▶
The programme settings and cleansers and disinfectants that must be used
are listed in the instructions for use of the washer disinfector.
▶
Directly after automated cleaning/disinfection, treat the medical device with
the KaVo care products and systems provided by KaVo.
7.2.3 Manual internal cleaning
Validated internal cleaning (removal of residual protein) can be accomplished
with KaVo CLEANspray and KaVo DRYspray.