T
M
▶ Press and hold down the "Motor anticlockwise direction" button, the "Tumbler
filler" button and the foot pedal simultaneously.
- A beeps sounds.
The intensive germ reduction switches to step 3.
The automatic rinsing process starts up.
▶ After the rinsing process, set down the instruments.
Note
The OXYGENAL concentration increases after the interruption of the intensive
germ reduction. The automatic rinsing process must run (approximately a 6 mi‐
nutes).
Intensive germ reduction during prolonged periods of non-use; 4
weeks
During extended periods of non-use, e.g. semester breaks at university, intensive
germ reduction must be carried out during the time of non-use. The germ reduc‐
tion fluid remains inside the system during the period of non-use.
Before the period of non-use
▶ Start the intensive germ reduction and stop it in step 4 after the signal tone
during the dwell time of 30 minutes.
▶ To prevent damage due to hoses getting kinked, remove hoses from the germ
reduction attachment and place them in the dentist support.
▶ Keep the germ reduction attachment in the spittoon bowl.
▶ Cover the treatment unit during extended periods of non-use.
After the period of non-use
▶ Turn on the central compressor.
▶ Switch on the water supply for the treatment unit after the system pressure is
achieved.
▶ Replace the instrument hoses of the dentist and assistant units in the places
provided in the germ reduction attachment.
▶ Turn the treatment unit on.
The unit automatically starts the remaining steps of the unfinished cycle.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene function of the water-conducting
system
28/35