2.6.2
Preparation of the tumbler holder and filler
CAUTION
Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
▶ Make sure that the unit is turned off before removing the tumbler filler or the
spittoon bowl.
▶ Always check the correct seating of tumbler filler and spittoon bowl before
turning on the unit.
Cleaning and disinfection of the tumbler holder and filler
Note
When assembling, make sure that the O-ring is present on the tumbler filler.
Note
Inspect the O-ring (Mat. no. 0.200.6128) for damage and exchange if necessary
to prevent water or oxidation damage.
▶ Remove the tumbler filler ① (Mat. no. 0.764.8541) by pulling upwards.
▶ Take off the tumbler holder ③ (Mat. no. 0.764.6153).
▶ Clean the tumbler holder and tumbler filler under running water.
▶ Wet the O-ring ② (Mat. no. 0.200.6128) with silicone grease
(Mat. no. 1.000.6403).
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit
19/35