2
Setup methods according to DIN EN ISO 17664
2.1
References
The reprocessing consists of the following basic steps:
▪ Cleaning and disinfection ( manually or automatically)
▪ Sterilisation
Note
Comply with national hygiene requirements, e.g. RKI guidelines.
CAUTION
Risk of injury during cleaning of the treatment unit.
Lack of instructions to the cleaning staff and lack of preparation of the treatment
unit can lead to the cleaning personnel sustaining injuries.
▶ Only trained professionals and instructed cleaning personnel may be present
in the treatment rooms.
▶ Position the chair for cleaning and turn the device off.
CAUTION
Product damage due to improper disinfection.
Malfunctions.
▶ Use the disinfectant as prescribed by the manufacturer.
▶ No spray disinfection, perform wipe disinfection only.
▶ Do not immerse product or product parts in liquids.
CAUTION
Improper disinfection may lead to health hazards and material damage.
Infection hazard to users and patients.
Damage to the sterile product.
▶ No hot air sterilisation, no chemical cold sterilisation, do not sterilise with eth‐
ylene oxide.
CAUTION
Health hazard and property damage due to non-compliance with servicing
schedule.
Infection hazard to users and patients.
Product damage.
▶ Comply with servicing schedule.
Note
The preparation methods for the instruments and motors are found in the sepa‐
rate instructions for use that accompany the instrument and motor packages.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.1 References
3/35