2.4.6
Servicing the turbine return air filter
Cleaning the turbine return air filter
Note
The liquid collector must be checked weekly.
▶ If liquid is in the filter , drain the liquid and replace the O-rings of the multiflex
couplings.
▶ If necessary, replace the turbine return air filter (Mat. no. 0.220.7926) and
clean the filter holder.
1058 TM/C/G
1058 S
Return air filter
Mat. no.
1058 TM/C: Filter
Mat. no. 0.220.7926
1058 S: Round plate 33
Mat. no. 0.200.3098
2.4.7
Reprocessing of the kit for physiological saline solution (op‐
tional)
The kit for physiological saline solution is optional.
Note
The silicone hoses of the kit must be sterilised after each patient.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
15/35