7 Maintenance
8. Workstation:
Visually evaluate and inspect the 3D image for visual defects such as artifacts using
the 3D imaging software.
NOTICE!
The device determines if the QC image is PASSED or FAILED according to measured
data, not based on the visible image quality.
NOTICE!
You should also compare the new QC image to the reference image taken during
the installation or the latest service. Doing this helps you ensure that the image quality has
remained constant.
9.
If the image fails or any visual defects are apparent, redo the QC program. If it fails again,
recalibrate the device according to
on page 73 or contact service.
10.
Detach the 3D QC phantom and the 3D QC phantom holder from the device.