4 User interfaces
3.
SETTINGS
Device settings:
Friendly name
Set a name for the device, which is shown next to the device
settings icon and on the imaging software.
3D Metal Artefact
Reduction
(3D devices only)
Enable or disable MAR, Metal Artefact Reduction. MAR is used to
reduce the effect of metals and other dense radiopaque objects
on the 3D image.
It is recommended to keep MAR activated.
Force device language
Change the language of the device GUI if needed.
It is recommended not to force the language of the GUI but
change the language profile of the workstation.
Default to Scout before 3D
(3D devices only)
Enable or disable automatic Scout image activation when a 3D
imaging modality is selected. It is recommended to have the
Scout image taking enabled.
NOTICE!
If the selection is set to "disabled", a Scout
image mode can still be activated manually in the 3D
imaging program selection view.
Volume level
Adjust the volume level (5-10) of the audible signals played by
the device.
NOTICE!
Does not affect exposure warning signal.
The default volume level setting is 8.
Button volume level
Adjust the volume level (0-10) of the audible signal played when
a button on the patient positioning panel is pressed.
The default volume level setting is 8.
Button sound
Select the sound that is played when
HOME
,
Test mode
or
Patient positioning lights
buttons on the patient positioning
panel are pressed.
36
ORTHOPANTOMOGRAPH
™
OP 3D