6 Using the device
13.
Adjust the patient's head position/orientation.
NOTICE!
To ensure optimal image quality, pay attention to the correct patient positioning.
LATERAL PROJECTION:
• The patient's Frankfort-Horizontal plane should be horizontal. Use the FH light as a guide.
PA PROJECTION:
• The patient's Frankfort-Horizontal plane should be horizontal.
NOTICE!
The horizontal light is not lit for PA projections.
PA REVERSE TOWNE PROJECTION:
• Turn the patient's head downwards until the angle between the patient’s Frankfort-
Horizontal plane and the horizontal light is around
30°
.
• The patient's mouth should be opened fully and pressed against the neck/upper chest.
NOTICE!
The horizontal light is not lit for PA projections.
ORTHOPANTOMOGRAPH
™
OP 3D
59