Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
86 / 100
▶
Press the "HYDROclean" button.
ð
The HYDROclean function is in the preparatory stage. The top display LED of
the button flashes quickly.
▶
Plug the suction hoses without connectors onto the hygiene centre.
▶
Press the "HYDROclean" button.
ð
The tube washing function starts.
The upper indicating diode will light up when the tube washing process is fin-
ished.
▶
Remove the suction hoses from the hose cleaning attachment and place
them in the appropriate location.
ð
The tube washing function is completed.
The process can be discontinued at any time by pressing the "HYDROclean"
button.
6.2.4 Reprocessing of the handpiece holder
NOTICE
Damage caused by inappropriate cleaning agents.
▶
Do not use acid-containing, sand-containing, and form-forming cleaning
agents.
▶
Lift the handpiece holder up and take it out.
▶
Check the drain.
Содержание DSEclinical 5198
Страница 1: ...Instructions for use DSEclinical Typ 5198...
Страница 98: ......
Страница 99: ......
Страница 100: ...1 007 8663 Fk 20191001 3 en...