5 Operation | 5.5 Operating the foot switch
56 / 100
5.5.9 Actuate blown air
▶
Remove the handpiece (such as turbine, motor) from the holder.
ð
The handpiece is active.
▶
Press the "Blown air" foot button.
ð
As long as the footswitch is pressed, blown air exits the removed handpiece
(not with the PiezoLED)
5.5.10 Preselect counterclockwise motor rotation
▶
Take motor from the holder.
ð
The handpiece is active.
▶
Slide the cross switch upward.
ð
The direction of motor rotation is reversed each time the cross-switch is ac-
tuated: counterclockwise rotation - clockwise rotation.
ð
The direction of motor rotation is displayed on the dentist element.
5.5.11 Adjusting the instrument light
▶
Slide the cross switch to the right. (spotlight function)
ð
Cold light "On" (even when Cold light: "Off" is preselected).
▶
Slide the cross switch to the left.
ð
Change the cold light status: "On/Off"
5.5.12 Charging the wireless foot control
The wireless foot control is operated by means of an installed rechargeable bat-
tery.
CAUTION
Risk of injury and material damage from incorrect use of the charger
for the wireless foot control.
Personal injuries, damage to the wireless foot control or the charger.
▶
Do not use the treatment unit during the charging process!
▶
Do not use the wireless foot control charger supplied to charge non-re-
chargeable batteries.
▶
Charge the wireless foot control with the charger supplied only.