The test body is produced automatically.
The production takes approx. 20 minutes to complete.
▶ Open the door of the KaVo ARCTICA Engine and take out the blank.
15.4
Measuring the test body
Note
Incorrect measuring values.
Incorrect measuring values can lead to collision inside the machine.
▶ Take note of the two markers when you define the measuring positions.
▶ Make sure that no errors are made during the measurement or transfer of the
measured values.
▶ Brake the test body off the blank.
▶ Clean the measuring surfaces of the outside micrometer
(Mat. no. 1.002.5839).
▶ Close the outside micrometer and make sure that the display shows 0.000.
Technician's instructions KaVo ARCTICA Engine
15 Calibrating the machine | 15.4 Measuring the test body
96/106