5.4
Replacing the pressure switch for tool chuck air
▶ Disconnect the connection cable from the right pressure switch for tool chuck
air (5 bar) ① (Mat. no. 1.008.4691).
▶ Dismount the pressure switch of the tool chuck air ① from the quick-plug
coupling using fork wrench SW13.
▶ To mount the pressure switch, proceed in reverse order.
▶ Secure the quick-plug coupling with Loctite 511.
The pressure switches are in pre-set condition at the time of delivery.
Technician's instructions KaVo ARCTICA Engine
5 Compressed air unit | 5.4 Replacing the pressure switch for tool chuck air
31/106