6
Troubleshooting
6.1
Exchanging the O-rings on the coupling on the supply hose
CAUTION
Hazard from improper care of the O-rings.
Malfunctions or complete failure of the medical device.
▶ Do not use Vaseline or other grease or oil.
Note
The O-ring on the coupling may only be lubricated with a cotton ball wet with KA‐
VO spray.
▶ Press the O-ring between your fingers to form a loop.
▶ Shove the O-ring to the front, and remove it.
▶ Insert new O-rings into the grooves.
6.2
Cleaning the spray nozzle
CAUTION
Hazard from insufficient cooling water.
Overheating of the medical device and damage to the tooth.
▶ Check the water channels and clean the water nozzles with nozzle needle
Mat. no. 1.003.8150 according to need.
▶ Clean the water passage in the water nozzle by using the nozzle needle
Mat. no. 1.003.8150.
Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643
SMARTtorque S619 C - 1.008.1642
6 Troubleshooting | 6.1 Exchanging the O-rings on the coupling on the supply hose
11/21