▶ Remove the cutter or grinder, place the INTRA spray nipple tip in the opening
and spray.
▶ Subsequently treat with the specified care agents and systems.
7.5.2
Care with KaVo SPRAYrotor
KaVo recommends servicing the product after each time it is used, i.e. after each
automatic cleaning and before each sterilisation.
▶ Place the product on the appropriate coupling of the KaVo SPRAYrotor and
cover it with a CLEANpac bag.
▶ Service the product.
See also: Instructions for use KaVo SPRAYrotor
7.5.3
Servicing with KaVo QUATTROcare PLUS
KaVo recommends servicing the product after each time it is used, i.e. after each
automatic cleaning and before each sterilisation.
▶ Remove the cutter or grinder.
▶ Servicing the product in QUATTROcare PLUS.
See also: Instructions for Use KaVo QUATTROcare PLUS 2124 A
Servicing of the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a week.
See also: Instructions for Use KaVo QUATTROcare PLUS 2124 A
Note
Handpieces must be taken off the service couplings before the chuck service can
be started and performed.
▶ Remove the service coupling chuck from the side hatch of the QUATTROcare
PLUS and attach it to coupling service point four, on the far right. A MULTIflex
adapter must be mounted there.
▶ Press the handpiece together with the guide bush of the chuck to be serviced
against the tip of the service coupling chuck.
▶ Press the button showing the chuck service symbol.
Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643
SMARTtorque S619 C - 1.008.1642
7 Reprocessing methods according to ISO 17664 | 7.5 Care products and systems - Servicing