5
Operation
Note
At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 min. without the instrument being attached; if there is a risk
of contamination from reflux or back suction, the system must be rinsed for 20-30
seconds.
5.1
Attaching the medical device
CAUTION
Damage from inaccurate coupling.
Inaccurate coupling (especially during the afterglow period) can destroy the high-
pressure lamp or the LED of a MULTIflex (LUX) / MULTIflex LED coupling or re‐
duce its service life.
▶ Check the seat of the turbine on the coupling by pulling on it.
WARNING
Release of the medical device during treatment.
A medical device that is not properly locked in place can release from the MUL‐
TIflex (LUX) / MULTIflex LED coupling during treatment.
▶ Before each use, check if the medical device is securely locked onto the
MULTIflex (LUX) / MULTIflex LED coupling by pulling on it.
▶ Mount the medical device accurately on the
MULTIflex (LUX) / MULTIflex LED coupling and push it backward until the
coupling audibly locks in the medical device.
▶ Pull on it to make sure that the medical device is securely affixed to the cou‐
pling.
5.2
Removing the medical device
▶ Hold the coupling tight, and pull the medical device off while twisting slightly.
5.3
Insert the milling cutter or diamond grinder
Note
Only use carbide cutters or diamond grinders that correspond to ISO 1797-1 type
3, are made of steel or hard metal and meet the following criteria:
- Shaft diameter: 1.59 to 1.60 mm
- Overall length SURGtorque S459 L: max. 25 mm
- Shaft clamping length SURGtorque S459 L: at least 11 mm
- Blade diameter: max. 2 mm
Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643
SMARTtorque S619 C - 1.008.1642
5 Operation | 5.1 Attaching the medical device
9/21