3.2
Technical Specifications
SURGtorque S459 L
Minimal drive pressure
2.1 bar (30 psi)
Drive pressure
2.1 to 4.2 bar (30 to 61 psi)
Air consumption
40 Nl/min
Idle Speed
350.000 to 450.000 rpm
Power
> 15 W
Recommended operating
pressure
2 to 3 N
Note
The air consumption, idle speed, and power values were determined at a drive
pressure of 2.8 bar (40 psi).
Can be attached to all MULTIflex (LUX) / MULTIflex LED couplings.
3.3
Transportation and storage conditions
CAUTION
It is hazardous to start up the medical device after it has been stored strongly
refrigerated.
This can cause the medical device to malfunction.
▶ Prior to start-up, very cold products must be heated to a temperature of 20°C
to 25°C (68°F to 77°F).
Temperature: -20°C to +70°C (-4°F to +158°F)
Relative humidity: 5% RH to 95% RH absence of condensation
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Protect from moisture
Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643
SMARTtorque S619 C - 1.008.1642
3 Product description | 3.2 Technical Specifications
6/21