92
Please charge in a well-ventilated area where the battery is not directly exposed to sunlight. Do
not charge the battery under rainfall, in morning dew or in an area where the humidity is high.
Do not cover the wheelchair with any waterproof cloth or other objects while charging. Don't
charge in temperatures lower than -10
℃
(14°F) or higher than 50
℃
(122°F) as the charger may not
work well and the battery may become damaged.
CAUTION
The charger is not guaranteed to be water resistant.
Valve Regulated Lead-Acid batteries do not have a memory function like cell phone batteries
have. Please get in the habit of recharging your VRLA batteries frequently.
New batteries MUST be fully charged prior to initial use of the powered wheelchair.
Always charge new batteries before initial use or battery life will be reduced.
As a general rule, recharge batteries as frequently as possible to assure the longest possible life
and to minimize required charging time. Plan to recharge them when use of the powered
wheelchair is not anticipated
WARNING
Keep your wheelchair away from flammable objects while charging, as it may lead to fire or
explosion of the battery.
When the batteryies are charged, they may produce hydrogen. Therefore, do not smoke while
charging and charge in a well-ventilated area.
Never connect or disconnect the plug (cord) with wet hands or if the plug (cord) is wet to avoid
possible electric shock.
Never attempt to recharge the batteries by attaching cables directly to the battery terminals or
clamps. DO NOT jump start your wheelchair!
Do NOT attempt to recharge the batteries while the powered wheelchair is in operation.
18.4. Charger
The battery charger supplied by Karma is for indoor use only, it must be protected from moisture and
external heat source.
CAUTION
Before charging, make sure to set the 115V 230V switch to match the local AC voltage. If the
switch is improperly set to a lower or higher voltage position, it will cause damage to the charger,
and, change the 115V 230V switch settings only when charger is disconnected from AC line
voltage.
WARNING
The fan inside the charger will start to work when you turn on the charger. (for 8A charger will
start to work as the temperature reaches higher than 50
℃
and stop running when the
temperature is lower than 50
℃
) If you find that the fan does not function properly, DO NOT use
the charger. Otherwise, the charger may overheat and start a fire.
Содержание Morgan Series
Страница 1: ...1 Powered Wheelchair User s Manual Morgan Series Morgan Sling Seat Morgan Seat Morgan KISS Seat...
Страница 2: ...2 release 04 2019...
Страница 9: ...9 3 Declaration of conformity Morgan Series...
Страница 85: ...85 Dahl Docking station Dahl docking station mounted on the floor Docking station and docking plate on chassis...
Страница 116: ...116...