MULTICAL® 402
64
Kamstrup A/S · Technical Description · 5512-742_M1_GB_01.2016
9.2
Cable influence
9.2.1
2-wire sensor set
Usually small and medium-size heat meters just need temperature sensors with relatively short cable lengths.
Thus, 2-wire sensor sets can be used with advantage.
Cable lengths and cross sections of the two sensors which are used as temperature sensor set for a heat meter
must always be identical, and must neither be shortened nor extended.
The limitations connected to the use of 2-wire sensor sets according to prEN 1434-2:2009 appear from the table
below.
Cable cross
section
[
mm
2
]
Pt100 sensors
Pt500 sensors
Max. cable length
[
m
]
Temperature
increase
[
K/m
]
Copper @ 20
°
C
Max. cable length
[
m
]
Temperature
increase
[
K/m
]
Copper @ 20
°
C
0.25
2.5
0.450
12.5
0.090
0.50
5.0
0.200
25.0
0.040
Table 10
Kamstrup supply Pt500 sensor sets with up to 10 m cable (2 x 0.25 mm
2
)
9.3
Installation
9.3.1
Electrical connection
The two paired 2-wire sensors must be mounted in terminals 5 and 6 (T1), and 7 and 8 (T2). The polarity of
temperature sensors T1 and T2 is unimportant.
See the position of the terminals below:
Terminal no.
Standard heat and cooling
measurement
T1
5-6
Sensor in forward pipe (red)
T2
7-8
Sensor in return pipe (blue)
Содержание MULTICAL 402
Страница 1: ...MULTICAL 402 Technical Description...
Страница 2: ...MULTICAL 402 2 Kamstrup A S Technical Description 5512 742_M1_GB_01 2016...
Страница 109: ...MULTICAL 402 Kamstrup A S Technical Description 5512 742_M1_GB_01 2016 109...