User Manual
C3-50
38
Alarm message
8.0
TROUBLESHOOTING
8.1 Alarm screen
In the case of an alarm the alarm message will appear.
8.2 Over temperature shut-off
Microprocessor controller has a special safety system which turns heaters off if temperature rises
for more than 5 °C above set temperature.
In case of an error on this system, a mechanical over temperature shut-off device will disable
heating.
If this happens the cause of overheating must be found and the over temperature shut-off device
must be manually reset.
The over temperature shut-off device is located behind the rear plate of the lower housing and is
manually reset by pressing the spring next to the thermostat setting shaft (See picture).
Over temperature shut-off device
(chamber heating)