Relative humidity control
For the purpose of humidity control unit is equipped with double flow system:
-
Dry air purging system which is used for maintaining low humidity, drives dry air in to the
chamber and pushes out the moist.
-
The saturated air purging system which is used for maintaining high humidity works
directly opposite of the dry air purging system and drives saturated air in to the chamber.
DRY AIR PURGING SYSTEM:
Compressed air supply
(max. 1MPa)
Air Supply pressure switch
(set to 0.25MPa)
Pressure regulator 1
(set to 0.4MPa)
Pressure regulator 2
(set to 0.02MPa)
Membrane Air Dryer
Membrane Air Dryer
Micro mist separator
Dew point indicator:
- Initial state: White, Pink
- Normal operating: Blue
- Decrease in performance: Brown, Black
Mist separator
Solenoid valve
Pressure regulator
(set to 0.02MPa)
SATURATED AIR PURGING
SYSTEM:
Located under the back bottom housing panel
Solenoid valve