User Manual
C3-50
34
7.3 Corrections setup scheme
1 - 4
15
11
17
16
ENTER ENTRY
CODE: 0207
CORRECTIONS SETUP
PT 100 CHAMBER
COR. Tc = 17.26°C
-40°C = 4.8°C
SPECIAL SETUP
COUNT DOWN TOLERANCE
IN TEMP. RANGE +- 1.0 °C
CORRECTIONS SETUP
RH PROBE
Rh = 30.0%
Offset Rh = 0.0%
SPECIAL SETUP
COUNT DOWN TOLERANCE
IN HUMID. RANGE +- 5%
°C
5-6
7
PID – AIR Rh Regul.
Kp= 6.0°C ( 6.0)
Ti= 150sec ( 150)
Td= 0 ( 0)
PID – TEC Chamber
Kp= 30.0°C ( 30.0)
Ti= 50sec ( 50)
Td= 0 ( 0)
SPECIAL SETUP
CICLE SERIAL
NUMBER = 13
SPECIAL SETUP
DISPLAY
CONTRAST: 5
14
DSPLAY
WORKING HOURS
AFTER SERVICE: 01:30
TOTAL: 01:30
SPECIAL SETUP
LANGUAGE: ENGLISH
10
3
13
SPECIAL SETUP
Tglass=30.0°C / 31.0°C
DOOR GLASS HEATING
Tset FACTOR :1.00x
SPECIAL SETUP
RS232 COMMUNICATION
BAUDRATE : 19200
8
9
PID – T Chamber
Kp= 2.0°C ( 2.0)
Ti= 150sec ( 150)
Td= 0 ( 0)
12