User Manual
C3-50
29
4. Cycling of steps
•
Choose how many times should the selected program sequences to be repeated.
5. Cycle from step
6. Cycle to step
•
With these two settings you select the program sequence that will be repeated.
For example:
you have “number of steps” set to 4, and you would like to repeat/cycle all steps
from 1 to 4. In this case you have to set the “cycle from step” to 4, and “cycle to step” to 1.
This way when the program is finished with step no. 4 it returns to step no. 1 and repeats the
process. The process will be repeated for so many times as you have set in the “cycling of
step” setting.
When you are using cycling options the main display will show:
Meaning:
S: 2
this number will indicate which step
(“number of steps” setting) of
the program is currently under execution.
7/20
20
–
number off all steps that need to be executed, 7
–
current step
in process.
Example:
“number of steps” is set to 5, cycling of steps is set to 3, “cycle from step” is 5,
“cycle to step” is set to 1. So the program shall do 5 steps and then repeat them for 3 times
(5 + 3 x 5 = 20). together we have 20 steps to execute. Currently is in step 7.
30:00
Countdown timer for the current step (step “duration time” setting).
Tc=38.3°C / 40°C
Current chamber temperature is 38.3°C, set chamber temp. 40°C
Rh= 62.2% / 75 %
Current chamber humidity is 62.2%, set humidity is 75 %.
ATTENTION:
“Cycling from step”
number must always be bigger than “
Cycling to step”
number!
S: 2, 7/20 30:00
Tc= 38.3°C / 40°C
Rh = 62.2 % / 75%