OWDMANUAL 1114
KAM CONTROLS, INC.
29
DISASSEMBLY
2.
3.
FIG. 5-2
SMA Connectors
RTD
Unscrew lid to round junction
box (FIG. 5-1) and ensure that
cables are somewhat slack. If
not, untie/loosen the cables prior
to pulling probe away from hous-
ing.
Slowly pull Probe away from
Housing to gain access to the
SMA connectors. Do not pull too
hard or too far as the wires can
be damaged. FIG. 5-2
Remove all (6) 8-32 Set Screws using the 5/64" Allen Wrench. Make sure that the wrench in fully inserted or the
Set Screws will strip. FIG. 5-1
1.
FIG. 5-1
8-32 Set Screws
Junction Box Lid
ANTENNA REPLACEMENT
1. Phillips Screwdriver Size 0
2. 1/16" Allen Wrench
3. 5/64" Allen Wrench
4. 5/16" Allen Wrench
1. Stainless Antenna
2. Titanium Antenna
3. (2) 2-004 O-rings
4. (2) 2-009 O-rings
5. Medium Strength Loctite
TOOLS REQUIRED
MATERIALS REQUIRED
Contact KAM CONTROLS
at +1 713 784 0000,
Fax to +1 713 784 0001,
or email [email protected].
Ask for Part Number: OWD ANT
M A I N T E N A N C E C O N T I N U E D