I N S TA L L AT I O N C O N T I N U E D
INPUTS
24V (–) IN
GND
24V (+) IN
24-30 VDC
DIG IN (–)
Pulse input, discrete input for different modes of operation (0 or 5 volt)
DIG IN (+)
AIN2
Density In – Not required on 0-100% models
OUTPUTS
4-20 mA (–)
Current output, source powered. Set to requisite percent water at factory.
4-20 mA (+)
AN OUT (–)
Can be 4-20 mA or analog voltage (used for local display or to pass density to PLC)
AN OUT (+)
DIG OUT (–)
Alarm or relay (digital contact closure)
DIG OUT (+)
INPUT/OUTPUT
RS232
Console port – communication interface for calibration, connection to PLC
RS485
Modbus interface
LED INDICATORS
D1
Power
SERIAL PORT CONNECTIONS
DB9 (female)
5
GND
3 RS232RX
2 RS232TX
All wiring and maintenance on the KAM OWD must be done in accordance with regional and classification re-
quirements. It is the user's responsibility to understand these requirements.
It is also recommended that the OWD be wired with flexible wiring/conduit with additional slack/length in the
wire to accommodate insertion, removal, and testing.
Operator's should take all possible precautions to avoid any moisture from entering the electronics enclosure.
The enclosure should not be left open in inclement weather or for long periods of time, especially during opera-
tion as condensation will accumulate. Lid should be tightly screwed shut, all conduits should be sealed and
secured in accordance with regional and classification requirements, and unused 3/4" NPT openings should be
seal with provided plug. Do not power wash the unit.