Dynamic gamma function
CL-S500 comes with our original function, dynamic gamma adjustment. By enabling this
function, the images on the screen are automatically identified to distinguish between
monochrome and color, and the respective images are displayed with the most suitable
grayscale. The default is “Disable”, but by using OSD operation can change it to “Enable”.
Note
This function is available only when the monitor has been calibrated to
DICOM GSDF.
Setting method 1
You can turn on and off DYNAMIC GAMMA by OSD.
1)
Activate “Operation Button Guide
①
” by pressing any one of operation buttons
(A–C).
2)
Press the “
MENU
” button.
3)
Select “
DYNAMIC GAMMA
” on MENU 1/2 by using the “▲” button or the “▼” button
and press the “
OK
” button.
Secondly, select “
Disable
” or “
Enable
” on OSD “DYNAMIC GAMMA” and press the
“
OK
” button.
Setting method 2
You can make function buttons (F1 or F2) available to switch the DYNAMIC GAMMA. By
doing so, you can easily switch from “Disable” to “Enable” by using a button.
The following steps tell you how to functionalize the “
F2
” button.
1)
Activate “Operation Button Guide
①
” by pressing any one of operation buttons
(A–C).
2)
Press the “
MENU
” button.
3)
Select “
FUNCTION
” on MENU 1/2 by using the “▲” button or the “▼” button and
press the “
OK
” button.
Secondly, select “
FUNCTION2
” on OSD “FUNCTION” and press the “
OK
” button.
4)
Select “
DYNAMIC GAMMA
” on OSD “FUNCTION2” and press the “
OK
” button.
5)
Activate “Operation Button Guide
①
” by pressing any one of operation buttons
(A–C)
.
6)
You can switch the status of “D-GAMMA” (Disable or Enable) by pressing the “
F2
”
button.
Also, you can check the status of “
D-GAMMA
” on OSD “INFORMATION 1/2”.