2 Device versions
12
2.5.3
Signal cable length and preamplifier
For devices in the remote mount design, the electrical connection between the transmitter and sensor is
made via a signal cable.
Transmitter housing
design
Single-compartment
housing
Ex-zone 2
Ex-zone 2 or
outside the Ex-area
Maximum signal cable
length
a
a
At a minimum conductivity for the measurement medium of ≥ 5 µS/cm
Without preamplifier
50 m (164 ft)
With preamplifier
200 m (656 ft)
Scope of delivery
5 m (16.4 ft) included
Signal cable, part no.
00648906
Transmitter housing
design
Single-compartment
housing
Ex-zone 1
Ex-zone 2 or
outside the Ex-area
Maximum signal cable
length
a
Without preamplifier
50 m (164 ft)
With preamplifier
-
Scope of delivery
5 m (16.4 ft) included
Signal cable, part no.
00648906
Transmitter housing
design
Dual-compartment
housing
Ex-zone 1
Maximum signal cable
length
a
Without preamplifier
10 m (164 ft)
With preamplifier
-
Scope of delivery
10 m (32.8 ft)
fixed installation
Signal cable, part no.
00648907
406018/2-0
406018/2-1
406019/2-0
406019/2-1
406018/2-0
406018/2-1
406019/2-0
406019/2-1