23
3 Mounting
3.4
Notes on using the device in areas with combustible dust
The device with dual-compartment transmitter housing is approved for use in potentially explosive areas
(gas and dust).
The Ex mark is provided on the nameplate.
3.4.1
Maximum admissible surface temperature
3.4.2
Minimum signal cable length
The signal cable in explosion protection areas may not be shorter than 5 m (16.4 ft).
DANGER – Explosion hazard!
Among other things, the dust explosion protection is ensured by the housing.
No changes may be made to the housing (e.g. removing or not using parts).
IMPORTANT (NOTE)!
The maximum surface temperature is applicable to dust layers of up to 5 mm (0.20 inches) in thickness.
This is to be used to determine the minimum admissible ignition and glow temperature for the dust at-
mosphere according to EN 60079.
The maximum admissible surface temperature must be reduced for thicker dust layers. The dust may be
conductive or non-conductive. EN 60079 ff. must be observed.
Model
Surface temperature
406012/1-1
70 °C (158 °F) to T
Medium
406013/1-1
406015/1-1
406016/1-1
406012/2-1
85 °C (185 °F) to T
Medium
406013/2-1
406015/2-1
406016/2-1
406018/2-1
70 °C (158 °F)
406019/2-1
Содержание 406012/1-1
Страница 2: ......
Страница 38: ...4 Ex relevant data for operation in zones 1 21 and 22 38...
Страница 50: ...6 Startup 50...
Страница 54: ...7 Maintenance 54...
Страница 56: ...8 Annex 56...
Страница 57: ......