19
3 Mounting
3
Mounting
3.1
Notes on opening and closing the housing
To install, start up and maintain the JUMO flowTRANS MAG devices, it is necessary to open the housing
of the sensor or transmitter.
Be sure to observe the following safety instructions before opening and after closing the housing for re-
starting:
3.1.1
Opening the case lid (dual-compartment housing)
1.
Observe chapter 3.1 "Notes on opening and closing the housing", page 19!
2. To open the housing, loosen the lid lock (1) by screwing in the Allen screw.
3. Remove the case lid.
4. Before closing the housing, make sure the case lid (O-ring) is connected properly.
5. After closing the case, secure the case lid by unscrewing the Allen screw to prevent the case from
being opened accidentally.
DANGER – Explosion hazard!
When the case lid is open, the Ex-protection becomes void.
Before opening the housing, de-energize all of the device's connecting cables and wait for at least 20
minutes.
DANGER – Explosion hazard!
Loosening the transmitter housing screws invalidates the Ex-protection. Before re-starting, tighten all
screws on the transmitter housing.
WARNING – Personal injuries!
When the housing is open, the EMC protection is limited and the protection against contact becomes
void.
Before opening the housing switch off the voltage supply.
WARNING – Impairment of the housing protection type!
If the seal (O-ring) is not connected properly or damaged, the housing protection type is impaired. Before
closing the case lid, check the seal (O-ring) for damage and replace if necessary.
When closing the case lid, make sure the seal is connected properly.
(1)
Lid lock (Allen screw)
Содержание 406012/1-1
Страница 2: ......
Страница 38: ...4 Ex relevant data for operation in zones 1 21 and 22 38...
Страница 50: ...6 Startup 50...
Страница 54: ...7 Maintenance 54...
Страница 56: ...8 Annex 56...
Страница 57: ......