49
11 The pH combination electrode
Problem cases
• with solutions containing protein, such as those used for measurements in
the medical and biological sector, KCl can result in the denaturation of the
protein
• coagulated coatings
• solutions that contain relatively high concentrations of silver ions
• ion-deficient solutions (de-ionized or ultra-pure water)
• solutions containing a solvent
If measurements are to be taken in these or similar solutions, we advise deter-
mining the pH value as quickly as possible so as to avoid unnecessary strain
on the combination electrode.
After the measurement, the combination electrode must be cleaned and
stored in either a KCl storage container or with a watering cap.
Substances that are deposited on the measuring membrane or the diaphragm
have an effect on the measurement and must be removed regularly. This can
be done, for example, using an automatic cleaning device.
11.1
pH combination electrodes for different areas of application
NOTE!
pH combination electrodes are wearing parts that need to be replaced after
chemical and mechanical exposure when the required values can no longer
be obtained, even after thorough cleaning.
When in use, it is important to remember that various substances (for exam-
ple, hydrofluoric acid (HF)) corrode glass in aqueous solutions. Chemicals
can react with the KCl electrolyte in the combination electrode and/or cause
a blockage of the diaphragm
.
Area of application
pH combination electrode -
version, type, part no.
Drinking water monitoring, aquariums, fish
breeding, lake and sea water, etc.
Type 201005/65-18-45-2000-76-120/837,
2 m fixed cable connection and BNC connector,
part no. 00417300
Generally aqueous media, beverage indus-
try, company laboratory, chemistry, etc.
Type 201030/76-10-07-40-1000-76-120/000, KCl-
filled, 1 m fixed cable connection and BNC connec-
tor
part no. 00300196
In the food industry for insertion measure-
ments in sausage, meat, cheese, fruit,
pasta products, and other semi-solid food,
soil measurements, small liquid samples,
etc.
Type 201030/60-15-08-40-1000-76-120/052,
insertion electrode, plastic version,
solid electrolyte gel, 1 m fixed cable connection and
BNC connector,
part no. 00448527
Industrial and municipal wastewater, elec-
troplating plants, paint shops, generally
contaminated media, etc.
Type 201030/51-18-07-40-2000-76-120/052, 837,
2 m fixed cable connection and BNC connector,
part no. 00532380
All manuals and user guides at all-guides.com
Содержание 202710
Страница 2: ...All manuals and user guides at all guides com ...
Страница 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 57: ...All manuals and user guides at all guides com ...