35
8 Device settings
Parameter "InP": selection of a measurand
✱
Select the desired measurand with the
and
keys:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
(4)
"CAL.P": selection of the buffer series, See page 37
(5)
"C.int": calibration timer, See page 37
(6)
"Unit t": temperature unit °C/°F, See page 38
(7)
"Auto HLD": automatic saving of a measured value, See page 38
(8)
"P.oFF": automatic device switch-off, See page 39
(9)
"Out": universal output, See page 39
(10)
"Adr.": universal output
Parameters
NOTE!
In the following list of parameters, the setting options for the default setting
are shown in bold in each case.
Menu/parameter
Setting option
Meaning
Arrow "rH"
Measuring (detecting) the rH value
Arrow "mV"
Measuring the redox potential
Arrow "mV
H
"
Measuring the redox potential with refe-
rence to the hydrogen system
Arrow "pH"
Measuring the pH value
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
All manuals and user guides at all-guides.com