29
6 Calibration of the pH combination electrode
✱
Use the
and
keys to set the value for the third buffer solution.
Rinsing
✱
Rinse the combination electrode with de-ionized (distilled) water and dry.
Calibration point 3
✱
Insert the combination electrode and the temperature probe into the third
buffer solution and gently stir.
✱
Press the
key when the measured value is stable (small triangle
appears on the display above the yellow "ok" label).
✱
Use the
and
keys to set the temperature value for the third buffer
solution and then save by pressing the
key.
Finally, the condition of the combination electrode is briefly displayed and then
the 3 point calibration is completed and the device returns to measurement
mode.
6.5
Display of the condition of the pH combination electrode
✱
Briefly press the
key.
The value on the main display of the device flashes,
"Pt.3" and a revolving symbol appear in the subsidi-
ary display of the device to indicate that the input of
the pH value for the third buffer solution is expected.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
The display changes so that the temperature value
(t.3) for the third buffer solution can be entered.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
NOTE!
After every calibration, the condition of the pH combination electrode is dis-
played for approximately 3 seconds.
The condition of the pH combination electrode is dis-
played for approximately 3 seconds.
100 %:
The combination electrode is in very good condition
90 to 30 %:
The condition of the combination electrode is adequate
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
All manuals and user guides at all-guides.com
Содержание 202710
Страница 2: ...All manuals and user guides at all guides com ...
Страница 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 57: ...All manuals and user guides at all guides com ...