6 Calibration of the pH combination electrode
30
6.6
Displaying specifications of the combination electrode
Displaying asymmetry
✱
Press the
key twice in an interval of approx. 1 second.
Displaying slope at 25 °C
✱
Briefly press the
key three times in an interval of approx. 1 second.
less than 30 %: The combination electrode is severely aged or contami-
nated.
If there is no improvement after cleaning (see chapter 11.2
"Cleaning", page 50)
and subsequent calibration, the com-
bination electrode must be replaced.
NOTE!
During the evaluation, both the asymmetry and the slope of the combination
electrode are considered.
The asymmetry of the pH combination electrode is
displayed for approximately 3 seconds.
Example: asymmetry -30 mV
Admissible range:
±55 mV
Optimal:
0 mV
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
NOTE!
Contamination can cause the asymmetry of the combination electrode to
deteriorate significantly.
The slope of the pH combination electrode is dis-
played for approximately 3 seconds.
Example: slope -52.9 mV/pH
Admissible range:
-62 to -45 mV/pH
Optimal:
-59.2 mV/pH
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
All manuals and user guides at all-guides.com
Содержание 202710
Страница 2: ...All manuals and user guides at all guides com ...
Страница 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 57: ...All manuals and user guides at all guides com ...